When does assembling a procedure pack turn you into a manufacturer under the MDR?
Is combining CE-marked devices into a procedure pack really a matter of packaging alone? It is easy to assume the answer is yes, because every device in the pack already bears a CE marking. Under Article 22 of the MDR, however, there are cases in which the pack is treated as a device in its own right and becomes subject to a conformity assessment under Article 52.
This month’s article sets out the three cases in which the role changes, what the statement must declare, and where the MDR draws the line between combining devices and acting as a manufacturer.
The myth: assembling CE-marked devices never changes your role
It is commonly assumed that combining devices that already bear the CE marking is a packaging activity and nothing more. Under the MDR, this assumption does not hold. Article 22(1) requires natural or legal persons who combine CE-marked devices with other devices or products “in a manner that is compatible with the intended purpose of the devices or other products and within the limits of use specified by their manufacturers, in order to place them on the market as a system or procedure pack” to draw up a statement. Article 22(4) names the cases in which the role changes.
The reality: how the pack is put together decides, not the act of assembling
Article 22 sets out how devices are combined and what the person who combines them declares. Article 22(4) then names the cases in which the system or procedure pack is “treated as a device in its own right” and is subject to the relevant conformity assessment procedure under Article 52, and the person then takes on the obligations of a manufacturer.
What counts as a system or procedure pack
A procedure pack is a combination of products packaged together and placed on the market to be used for a specific medical purpose (Article 2(10)). A system is a combination of products, packaged together or not, that are intended to be interconnected or combined to achieve a specific medical purpose (Article 2(11)). Besides other CE-marked devices, a pack can include CE-marked in vitro diagnostic medical devices and other products that conform to the legislation applicable to them, where they are used within a medical procedure or their presence is otherwise justified (Article 22(1)).
What the statement must declare
The person who combines the devices declares that (Article 22(2)):
the mutual compatibility of the devices and, if applicable, other products was verified in accordance with the manufacturers’ instructions, and the activities were carried out in accordance with those instructions
the system or procedure pack was packaged and relevant information was supplied to users, incorporating the information from the manufacturers
the activity of combining the devices was subject to appropriate methods of internal monitoring, verification and validation
Sterilisation
Whoever sterilises a system or procedure pack to place it on the market applies, at their choice, one of the procedures in Annex IX or the procedure in Part A of Annex XI. The involvement of the notified body is limited to the aspects of ensuring sterility until the sterile packaging is opened or damaged. A statement declares that sterilisation was carried out in accordance with the manufacturer’s instructions of the individual components (Article 22(3)).
When the obligations of a manufacturer apply
Article 22(4) names three cases. The pack is treated as a device in its own right, and the person takes on the obligations of a manufacturer, where:
the pack incorporates devices that do not bear the CE marking
the chosen combination of devices is not compatible in view of their original intended purpose
sterilisation has not been carried out in accordance with the manufacturer’s instructions
Labelling and records
A system or procedure pack does not bear an additional CE marking. It bears the name, registered trade name or registered trade mark of the person who put it together and the address at which that person can be contacted, and it is accompanied by the information referred to in Section 23 of Annex I. The statement is kept at the disposal of the competent authorities for the period that applies under Article 10(8) to the devices that have been combined, which is at least 10 years after the last device has been placed on the market and at least 15 years for implantable devices. Where those periods differ, the longest period applies (Article 22(5)).
KEY TAKEAWAY
The practical takeaway for you
Combining CE-marked devices into a system or procedure pack falls under Article 22, and Article 22(4) names the cases in which the person who combines them takes on the role of a manufacturer. The person who combines the devices verifies their compatibility in accordance with the manufacturers’ instructions, packages the pack with the relevant information for users, validates the combining activity and draws up the statement.
If a device does not bear the CE marking, if the combination is not compatible in view of the original intended purpose, or if sterilisation does not follow the manufacturer’s instructions, the pack is treated as a device in its own right, the conformity assessment procedure under Article 52 applies, and the person takes on the obligations of a manufacturer. The role depends on how the pack is put together, and a CE marking on every device in it is not enough on its own.